Surface Characterization of Retrieved Metal-On-Metal Total Hip Implants from Patients with Adverse Reaction to Metal Debris

dc.contributor.authorBurbano-Salazar, Mariaen_US
dc.contributor.authorRussell, R.en_US
dc.contributor.authorHuo, M.en_US
dc.contributor.authorWelch, R.en_US
dc.contributor.authorRoy, Dianaen_US
dc.contributor.authorRodrigues, Danieli C.en_US
dc.date.accessioned2014-10-02T21:59:33Z
dc.date.available2014-10-02T21:59:33Z
dc.date.created2014-03en_US
dc.description.abstractThe use of metal-on-metal (MoM) total hip implants has decreased recently due to reports of high failure rates and adverse local tissue reaction (ALTR). It has been hypothesized that wear metal debris released from CoCr bearing surfaces may provoke delayed hypersensitivity reactions. The goal of this study is to evaluate the microscopic bearing surface characteristics of implants revised due to evidence of ALTR. The bearing surface of each head and cup was analyzed using multiple microscopy techniques for characterization of the surface features. The presence of severe mechanical scratching was a common characteristic found in all of the implants evaluated. Mechanical factors seemed to be the prevalent failure mode related to the appearance of ALTR with this particular set of retrieved implants.en_US
dc.identifier.citationBurbano, Maria, Robert Russell, Michael Huo, Robert Welch, et al. 2014. "Surface characterization of retrieved metal-on-metal total hip implants from patients with adverse reaction to metal debris." Materials 7(3): 1866-1879.en_US
dc.identifier.issn1996-1944en_US
dc.identifier.issue3en_US
dc.identifier.startpage1866en_US
dc.identifier.urihttp://hdl.handle.net/10735.1/4063
dc.identifier.volume7en_US
dc.language.isoenen_US
dc.publisherMDPI AGen_US
dc.relation.urihttp://dx.doi.org/10.3390/ma7031866en_US
dc.rightsCC BY 3.0 (Attribution)en_US
dc.rights.holder©2014 The Authorsen_US
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/en_US
dc.source.journalMaterialsen_US
dc.subjectHip Prosthesisen_US
dc.subjectCorrosionen_US
dc.subjectOptical microscopyen_US
dc.subjectMetal-on-Metal Joint Prosthesesen_US
dc.titleSurface Characterization of Retrieved Metal-On-Metal Total Hip Implants from Patients with Adverse Reaction to Metal Debrisen_US